Product
Setaflash Series 8 ActiveCool Flash Point Tester – Gas Ignitor and Corrosion Resisting Cup - 82160-2
The Setaflash Series 8 ActiveCool, closed cup flash point tester provides fast and reliable flash point results with a wide temperature range of -30 to 135 °C with no external cooling required.
With the Series 8, a test takes less than 2 minutes with just a 2 or 4 ml sample size. A simple user interface with colour digital display and touchscreen icons guides you through the straightforward testing process. For easy record keeping the instrument stores 100,000 test results which can be saved and transferred via a USB port.
Designed for precise and accurate tests, barometric pressure is measured by an internal barometer and correction is automatically applied to the test result. Unknown samples can be tested using ramp mode.
Learn more about flash point testing and which instrument you should be using >
Why flash point autosamplers do not comply with flash point test methods >
Setaflash Series 8 ActiveCool Flash Point Tester – Gas Ignitor and Corrosion Resisting Cup - 82160-2
Features
- Simple test procedure
- Compact, rugged design
- Gas ignition
- Automated flash detection
- Ramp mode for unknown samples
- Automatic barometric pressure correction
- Peltier heating and cooling
- Fire detection (alarm and relay)
- USB port and result storage
Information
Details
Ease of use:All functions on the Series 8 including test parameters, instrument status and test results are accessed via the touchscreen display. An audible prompt sounds when a flash is achieved and an automatic flash detector enhances repeatability.
Safety:
Peltier cells prevent the instrument from overheating, whilst a fire sensor immediately stops the test and cools the sample cup if a fire is detected.
Calibration and verification:
Calibration of the Series 8 can be performed in the field, eliminating the costs and time associated with sending the instrument to a service centre.
Service and maintenance:
The rugged design ensures minimal operational maintenance is needed.
Supplied with: 2 ml syringe, 99878-3 CRM, Allen keys, mains lead and instruction manual.
Details
ASTM D3278 | Standard Test Methods for Flash Point of Liquids by Small Scale Closed-Cup Apparatus - paints, enamels, lacquers, varnishes, and related products |
---|---|
ASTM E502 | Standard Test Method for Selection and Use of ASTM Standards for the Determination of Flash Point of Chemicals by Closed Cup Methods |
ISO 3679 | Determination of flash point. Rapid equilibrium closed cup method |
ASTM D3828 | Standard Test Methods for Flash Point by Small Scale Closed Tester |
IP 523 | Determination of flash point. Rapid equilibrium closed cup method |
ASTM D7236 | Flash Point by Small Scale Closed Cup Ramp Method |
IP 534 | Flash Point by Small Scale Closed Cup Ramp Method |
ASTM D1655 | Standard Specification for Aviation Turbine Fuels |
IP 602 | Determination of sustained combustibility of liquids |
CLP Regulations EC No 1272/2008 | European Regulation (EC) No 1272/2008 on classification, labelling and packaging of substances and mixtures |
BS EN ISO 3679 & BS 2000-523 | Determination of flash no-flash and flash point. Rapid equilibrium closed cup method. |
ASTM D8174 | Standard Test Method for Finite Flash Point Determination of Liquid Wastes by Small-Scale Closed Cup Tester |
EPA 1020 B | Standard test methods for flash point by Setaflash (small-scale) Closed-Cup apparatus |
SW-846 1020C | Standard Test Methods for Flash Point by Setaflash (Small Scale) Closed-Cup Apparatus |
Details
Temperature range | -30 to 135 °C |
---|---|
Temperature selection | °C or °F |
Ignitor | Gas |
Heating/cooling method | Peltier cell |
Temperature ramp mode | Automatic or custom settings (ramp rate up to 10 °C/min) |
Sample cup material | Stainless Steel |
Pressure correction | Automatic |
Data storage | 12 GB |
Voltage | 110 to 250 Vac, 50/60 Hz |
Size/Weight | 300 x 340 x 380cm / 8 kg |
CCCN Code | Tariff 90268020 |
Methods
Industries
Spares and accessories for Setaflash Series 8 ActiveCool Flash Point Tester – Gas Ignitor and Corrosion Resisting Cup - 82160-2
Printer Paper (Pack of 20) - 80602-001
View accessoryFlash Detector for Series 8 - 82000-201
View accessorySample Well O-ring, Viton (Pack of 5) - 13740-004
View accessorySyringe 4 ml (Small Scale Flash Point Instruments) - 13770-311
View accessorySyringe 2 ml (Small Scale Flash Point Instruments) - 13740-009
View accessoryDownload product profile
To view our full range of products, download our latest Product Profile here.
Want to know more?
View our full range of products, or simply get in contact with us for more information.